I have level data from a non-vented transducer in a water body and barometric pressure data from the same location. My elevation is 500 meters, how do I determine the true depth?
The true depth can be calculated by subtracting the actual atmospheric air pressure from absolute referenced level data.
Typically a non-vented depth transducer will have an absolute range, but this should be verified by checking the reading at zero depth, which should be around 10 mH20 or 1 bar, which would signify that the transducer does indeed have an absolute reference.
Featured non-vented reference hydrostatic level measurement products
DMP331P Hygienic Flush Pressure Transmitter - Sanitary pressure sensor wth exposed diaphragm with no crevices for high temperatures up to 300 Degrees Celsius in ranges from 100mb up to 40 bar gauge or absolute
IMCL Submersible Absolute Reference Depth Sensor - Absolute reference hydrostatic depth sensor with an IP68 sealed electrical interface, SS316L, Duplex or PVC housing and a mV, voltage or current output.
If the reading is close to zero, it will most likely be a sealed gauge reference type instead, this means that it will have a fixed zero offset that will typically correspond 1 bar, which will need to be determined precisely by subtracting the zero depth reading from the actual atmospheric air pressure and then adding the result to all non-vented depth readings, e.g. if Zero depth = -50 mbar, Atmospheric pressure = 980 mbar, then 1030 mbar should be added to all the depth data to convert it to absolute readings.
If the barometric pressure is the actual pressure at that location without any meteorological adjustment for elevation you can simply subtract that reading from the absolute referenced level reading, but be careful to ensure all calculations are carried out in the same pressure units.
If the barometric pressure was obtained from a meteorological service or barometer it is most likely to be referenced to sea level and you will need to compensate it for the 500m elevation by using a barometric formula to calculate corrected pressure for elevation.
Featured barometric pressure measurement products
Barometric 550 to 1150 hPa abs high resolution pressure transmitter - High precision barometric pressure transmitter for measuring atmospheric pressure from 550 hPa to 1150 hPa absolute.
Air flow normalising calculation barometric pressure transmitter - This pressure transmitter is used to measure barometric pressure for the purpose of normalising air flow calculations.
Related Help Guides
- Difference between vented and non-vented water level measurement
- Atmospheric pressure compensation for hydrostatic level sensors
- Using absolute pressure sensors to measure hydrostatic level
- Simulating 8000 foot altitude with a pressure gauge
- Measuring barometric pressure using a dp sensor
Related Technical Terms
- at – Technical Atmosphere Pressure Unit
- atm – Standard Atmosphere Pressure Unit
- inHg – Inches of Mercury at 0 degrees C Pressure Unit
- MSL – Mean Sea Level
Related Online Tools
- Gauge + Barometric to Absolute Pressure Calculator
- Suction Pressure to Vacuum Calculator
- US Standard Atmosphere Altitude and Pressure Calculator
- ICAO Standard Atmosphere Altitude and Pressure Calculator
- Elevation, Station (QFE) and Sea Level (QNH) Pressure Calculator
Related Product and Application Guides
Related Application Questions and Answers
- Ex rated ambient pressure sensor with 800-1100 mbar range
- 1013 to 1400 mbar barometric pressure transmitter & readout
- Wind turbine USB PC logging pressure & temperature sensor
- Gas sensors testing barometric pressure sensor with RS485 Modbus RTU interface
- Atmospheric range for 4-20mA pressure sensor
- 5 to 20 psi absolute ambient 4 to 20 ma pressure transmitter
Contact us about this Converting non-vented depth readings & barometric pressure to true depth page to request more information, or to discuss your application requirements.