Symptons and Causes of Damage to Pressure Transducer Diaphragms

It is possible to view part of the diaphragm on most pressure transducers and some which are completely exposed such as flush diaphragms are particular vulnerable to mechanical damage.

The first sign of damage may not actually be that detectable to the eye, instead it may show up as a small zero setting offset on the output signal. In some cases the damage maybe so severe that the diaphragm has completely ruptured compromising pressure containment or it is touching an internal component causing mechanical interference or an electrical short/open circuit.

Some of the signs which may point to a damaged diaphragm:

  • Traces of oil leaking from the pressure port
  • Decaying pressure output signal over time
  • Sudden change in output signal at certain pressures
  • Complete loss of output signal
  • Output signal stays at fixed value when pressure changes
  • Large change in zero offset reading
  • Significant change in reading characteristic or accuracy over the pressure range of the device

NB – Above symptoms may also be attributed to other modes of failure.

Overpressure protected products
  • PD39X High Differential Pressure Transmitter PD39X High Differential Pressure Transmitter
  • DPT200 Pressurised Tank Level Differential Pressure Transmitter
  • Liquid compatible transmitter with 15 psi DP range and protected on either side to 300 psi
  • DPT200 High One Side Overload Differential Pressure Sensor

Request info on overpressure protected products for your application.

Once a diaphragm has been damaged it is generally not possible to repair it. Some manufacturers incorporate the ability to replace a sensor module into the design on more expensive devices but for most products it is not economically viable to repair and therefore a replacement is the only option if the performance is deteriorating. Even if the performance change is acceptable the mechanical integrity of the diaphragm maybe compromised so much that a premature failure of the diaphragm is more likely to occur, shortening the service life.

A pressure transducer diaphragm is designed for many cycles of pressure in order for it to have a long service life and during the design and development phase it will be rigorously tested for mechanical integrity. A diaphragm failure, particularly at an earlier time in the service life, is more likely to be caused by an event which exceeds the specification limits of the transducer.

Some possible causes of diaphragm damage:

  • Sustained over-pressure
  • Cavitation from dynamic changes in liquid flow
  • Pressure spikes/surges
  • High frequency dynamic fatigue
  • Point loading of diaphragm with screw-drivers, pens or any pointed tool
  • Corrosion from incompatible media
  • High ‘g’ impact
  • Vibration at diaphragm resonant frequency
  • Expansion due to a change in media state from a liquid to a solid e.g. water and ice

A pressure transducer diaphragm is designed to withstand much of what an industrial environment will throw at it. There are some hazards though that should be avoided to protect what is after all a sensitive measurement instrument.

Overpressure protected products
  • DPT200 High One Side Overload Differential Pressure Sensor
  • 1psi wet/wet differential pressure transmitter for 2000 psi line pressure
  • Measure filter dp to 20mbar & withstand 1bar cleaning pulse
  • PD39X High Differential Pressure Transmitter PD39X Twin Absolute Digital DP Sensor

Request info on overpressure protected products for your application.

Primary Sidebar

Pressure Sensor Specifier

Specify a pressure sensor
Select your pressure sensor requirements

Product Selection Guides

  • Differential Pressure Sensors
  • Digital Pressure Gauges
  • Digital Pressure Sensors
  • Digital Readouts
  • Hydrostatic Liquid Level Sensors
  • Millivolt Output Pressure Transducers
  • Pressure Transducers
  • Pressure Transmitters
  • Voltage Output Pressure Transducers

Application Selection Guides

  • Absolute Pressure
  • Dual Temperature and Pressure
  • Flush Diaphragm
  • High Frequency Response
  • Leak Testing
  • Low Pressure Range
  • Oxygen Service
  • Submersible
  • Vacuum

Help Guides

  • Using absolute pressure sensors to measure hydrostatic level
  • Measuring vacuum with negative gauge or absolute ranges
  • Measuring the difference in air pressure between rooms
  • What does negative and positive gauge pressure mean
  • Measuring the Pressure of Hot Media
  • Determining water tank volume using hydrostatic pressure
  • How to get a 10 volt signal from a 4-20mA output pressure sensor

Product Enquiry

Send us your enquiry for a product associated with this Symptons and Causes of Damage to Pressure Transducer Diaphragms page, and get assistance with selecting a product for your application.

Email Your Enquiry

enquiries@sensorsone.com

© 2022 SensorsONE Ltd, all rights reserved

  • Home
  • Products
  • Tools
  • Legal
  • Contact