Is it possible to have a vacuum pressure measurement device with a -20 to +50 psi gauge pressure range, since I’ve never seen a pressure instrument with a range greater than -15 psi?
The majority of negative gauge vacuum range instruments are scaled to -14.5 psi or -15 psi to measure over the full vacuum range. The full vacuum range will vary because the surrounding atmospheric pressure which defines the vacuum threshold is constantly changing. However you can’t have a constantly changing negative gauge pressure range for the measurement device, so a convenient approximation of atmospheric pressure of 1 bar or 14.5 psi is used to scale and calibrate the full vacuum range of pressure measurement instruments.
- 2barg, 10barg range wireless pressure sensing kit for production monitoring use
- Vacuum logger to measure 0 to -1 bar suction or 0 to 1 bar absolute
- 10 psi vacuum air and 50 psi water pressure gauge and data logger
- HART coms compatible vacuum transmitter for -1 to 0 bar g
Since pressure measurement instruments are generally not capable of measuring high vacuum to high level of accuracy, they are only used to measure over the coarse vacuum range, so rarely are the instruments relied on to measure full vacuum precisely. The reason why this is important to the discussion is because it helps explain why it is not considered necessary to set the vacuum range to a greater negative pressure such as -20 psi.
If you consider the fact that atmospheric pressure is often greater than 1000 mbar 14.5 psi, then it is sensible to conclude that a vacuum range of -14.5 psi gauge will not cover the full vacuum range. So why is the negative pressure range not greater than 14.5 psi in this case? It is probably because of a few factors, and one of those would be that the extended vacuum range would hardly ever be needed because the instrument was never intended to measure precisely at high vacuum. Another reason would be that barometric pressure at sea level, although constantly changing, stays within a fairly narrow band rarely going as high as 1050 mbar.
So that explains why we tend to only see -15 psi ranges offered by instrument manufacturers, but is it actually feasible to have a greater range of say -20 psi g?
- Marine and IS approved, -1 to 3 mWG compound range pressure transmitter
- Under and Over pressure monitoring -10…+200 mb pressure transmitter
- Dredging vacuum pressure Transmitter
- 2barg, 10barg range wireless pressure sensing kit for production monitoring use
In order to scale a negative gauge range sensor or instrument it is necessary to apply a known pressure in order to calibrate the device. Zero pressure is very easy to set, since you only have to vent the measurement port to ambient air pressure to check this and set the reading. However setting -20 psi will be impossible unless the atmospheric pressure happens to be 20 psi absolute or you are using a environmental test chamber that can pressurise to 20 psi absolute.
This is when you call upon the extrapolation method to determine the theoretical reading at -20 psi gauge. The simplest approach is to measure as low as possible, e.g. -14 psi and then make the assumption that it is 70% of what would be read at -20 psi. If the characteristics of the measurement device are understood, then it maybe possible to apply a non-linear correction to the calibration that would produce a better estimate of the reading at -20 psi.
So although it is probably unnecessary to have a -20 psi gauge range for the majority of negative gauge pressure range applications, it is certainly possible to manufacture and calibrate a negative 20 psi gauge range.
- ATEX approved negative 10 mbar vacuum pressure transmitter
- -1 to +2 bar g compound pressure transducer with 5Vdc output
- Minus 15 psi gauge intrinsically safe pressure transmitter
- -760mmHg to +20 kg/cm² pressure transmitter with HART coms
Related Help Guides
- Measuring vacuum as a negative gauge pressure using a dp sensor
- Measuring vacuum with negative gauge or absolute ranges
- What is the difference between vacuum and absolute pressure
- Can you measure vacuum using a gauge pressure range
- What does negative and positive gauge pressure mean
- Measuring negative pressure using a positive differential pressure range