• Skip to primary navigation
  • Skip to main content
  • Skip to primary sidebar
SensorsONE

SensorsONE

Measurement Instrumentation Products

  • Products
  • Tools
  • Contact

Choosing calibrator for pressure transmitters

XMD Process Plant DP Cell Differential Pressure TransmitterA guide on what to consider when choosing pressure calibration equipment for calibrating pressure transmitters.

Contents

  • Required accuracy
  • Connecting device under test
  • Generating test pressures
  • Measuring 4-20mA output
  • Overall calibration accuracy
  • Calibrator accuracy stability

Required accuracy

First you must determine the highest overall accuracy you will need to achieve for the pressure transmitters you will be calibrating.  This does not necessarily have to be the manufacturer’s specification since very often the pressure transmitter accuracy is much better than the required system accuracy.

Since the required accuracy is not a parameter that is usually compromised, eliminating portable pressure calibrators without the required accuracy has to be the first step in narrowing the choices.

Ideally a 10 to 1 ratio is recommended for the difference in accuracy between the pressure transmitter and the pressure calibrator, so that no significant extra uncertainty is introduced into the measurements by the pressure calibrator.

However this is not always practical especially when calibrating high accuracy smart pressure transmitters and it is not unusual for a pressure calibrator with a much closer ratio to be used instead. In these cases the uncertainty of the pressure calibration equipment should also be considered in the overall system uncertainty budget.

Pressure calibrator products
  • 30bar pressure calibration handpump kit & 0.2% acc test gauge
  • 20 bar pressure calibrator with accuracy better than 0.5% down to 200 mbar
  • Calibration test pump and gauges for ranges 1000 mbar to 10 bar gauge Calibration test pump and gauges for ranges 1000 mbar to 10 bar gauge
  • PGS700 High Pressure Calibration Hand Pump

Connecting device under test

To connect the pressure calibrator you will need to isolate the pressure transmitter under test from the system so that only a small volume will need to be pressurised. Typically a 2 way or a 3 way valve is included in the pressure transmitter installation so that it can be isolated and removed for test or connected to a pressure calibrator in-situ.

Generating test pressures

The next consideration is how you are going to generate pressures. It is possible to source pneumatic calibration hand pumps which generate pressures up to 40 bar, however if the volume is too great you may struggle to achieve the manufacturer’s maximum pressure rating.

PGS40-700-1000A hydraulic pump will be needed to achieve calibration pressures above 40 bar.

In order to precisely set pressures it is important that a volume adjuster is incorporated into the calibration hand pump. This will enable each calibration set-point to be applied in one direction without over-shoot so that no pressure hysteresis is introduced into the calibration results.

To save on the number of pieces of test equipment which are carried around, some pressure calibrators include integral hand pumps and volume adjusters. These pressure calibrators are very convenient for portable use but when it comes to sending them away for servicing or switching pressure ranges there is no way to separate the pump from the pressure calibrator.

Measuring 4-20mA output

ADT672To measure the output signal of a pressure transmitter you can either break into the system current loop to measure the 4 to 20mA signal using a current meter or you can disconnect the pressure transmitter entirely and use a current meter that also has a loop supply to provide power to the pressure transmitter.

Again you can cut down on the number of test equipment by selecting a pressure calibrator that has an integral current meter with a loop power supply. This is particularly useful if the pressure calibrator includes a data logging function, because the calibration results can be stored in the memory for processing and producing a calibration certificate without the need to write down the calibration results.

Overall calibration accuracy

It is often overlooked but the overall accuracy of pressure calibration should also include the accuracy of the current meter. Another calibration error which should be investigated is the temperature error associated with both the pressure calibrator and the current meter. Some manufacturers will quote an overall accuracy for their pressure calibrator over a typical ambient temperature range, while others will quote the temperature errors separately. It is therefore important to examine a little more closely what makes up the accuracy statement on the data sheet when comparing one pressure calibrator with another.

Calibrator accuracy stability

Lastly it is important to point out that no portable pressure calibrator is perfectly stable and most manufacturers will include a stability statement which is either quoted separately or combined with the overall accuracy statement to provide a 3, 6 or 12 month figure. From this combined accuracy and stability figure you will be able to determine the re-calibration period for the portable pressure calibrator. Since the performance of individual instruments can vary, the re-calibration period can be widened if before and after calibration data shows that the pressure calibrator is holding its accuracy over time and this will save on re-calibration costs.

Pressure transmitter products
  • DMK351 Liquid and Gas Resistant Intrinsically Safe Low Range Gauge Pressure Sensor
  • HART® Pressure Transmitters
  • Differential Pressure Transmitters
  • DMP331 Precision Pressure Transmitter

Primary Sidebar

Pressure Sensor Specifier

Specify a pressure sensor

Select your pressure sensor requirements

Wirelessly Measure & Log 4-20mA

Use this battery powered wireless 4-20mA wireless transmitter to transmit readings to a receiver connected to USB, RS485, RS232, cellular network or analog output

Use this battery powered 4-20mA wireless transmitter to transmit readings to a receiver connected to USB, RS485, RS232, cellular network or analog output

Help Guides

  • Measuring negative pressure using a positive differential pressure range
  • Pressure Sensor Accuracy Specifications
  • Process Connections
  • Determining calibration error of Bourdon tube pressure gauge
  • Measurement Accuracy
  • Supply voltage and load resistance considerations for pressure transmitters
  • Measuring vacuum with negative gauge or absolute ranges
  • What is the difference between gauge and absolute pressure measurement
  • Measuring the difference in air pressure between rooms
  • How to Connect a 4-20mA Current Loop Pressure Transmitter
  • Advantages and Disadvantages of using a 4 to 20 mA Signal
  • Determining water tank volume using hydrostatic pressure
  • How do you measure the volume of a liquid using pressure
  • What is the difference between vacuum and absolute pressure
  • What does negative and positive gauge pressure mean

Featured Product Types

  • USB Pressure Sensors
  • Sea, Brackish or Salt Water Submersible IP68 Depth & Level Sensors
  • Chemical Resistant Hydrostatic Liquid Level & Pressure Sensors
  • RS232 Serial Interface Pressure Sensors
  • Differential Pressure Transmitters
  • Vacuum Pressure Transmitters

Product Enquiry

Send us your enquiry for a product associated with this Choosing calibrator for pressure transmitters page, and get assistance with selecting a product for your application.

Email Your Enquiry

enquiries@sensorsone.com

© 2023 SensorsONE Ltd, all rights reserved

  • Home
  • Products
  • Tools
  • Legal
  • Contact