Most air hand pumps that are designed to be used as pressure calibrators include a coarse pressure generator and a fine pressure adjuster. The effectiveness of both is dependent on the test volume connected to the pump.
The coarse or main pressure generator is typically a piston and cylinder arrangement which is actuated either by a set of calipers, a plunger, or a lever. The coarse change in pressure is applied by compressing the air to a high level inside a small cylinder which is then released to the lower pressure test volume increasing the pressure upon each cycle of the pump.
The fine pressure adjuster is a thumb-screw mechanism that incrementally reduces or increases the closed test volume which in turn changes the stored pressure by an equivalent proportion in accordance with Boyle’s Law.
Featured calibration hand pump products
The number of coarse pump cycles or turns of the fine volume adjuster required to reach a desired pressure, depends on the test volume connected to the pump. For example If the test volume is doubled
the number of cycles or turns will be doubled to achieve the same pressure change.
When testing a high pressure with a pneumatic hand pump it is essential to keep the test volume as small as possible. Also, winding out the fine adjustment screw as far as possible before applying any pressure will provide you with a convenient pressure booster when you or the pump runs out of steam at high pressures.
It is very difficult to set very low pressures with a small test volume due to the relatively large capacity of most calibration hand pumps. Increasing the test volume can help reduce the sensitivity of the coarse pressure pump, but alternatively using the fine volume adjuster for applying and incrementing the test pressure may produce the same results.
Featured calibration hand pump products
Related Help Guides
- Reduce calibration costs of analogue pressure gauges
- Choosing calibrator for pressure transmitters
- How does the accuracy of pressure measurement devices change over time
- Checking the LHR error of a 0-5 Vdc output pressure transducer
Related Technical Terms
Related Online Tools
- Pressure Transmitter 4-20mA Current Output Calculator
- Pressure Sensor Calculator
- DP Flow Transmitter Output Calculator
- Pressure Transducer Millivolt (mV) Output Calculator
- Pressure Transducer 0-10V Voltage Output Calculator
- Pressure Transducer 0-5V Voltage Output Calculator
- Pressure Transducer 1-5V Voltage Output Calculator
- Pressure Transducer 0.5-4.5V Voltage Output Calculator
- Pressure Sensing Errors Calculator
- Pressure Transmitter 0-20mA Current Output Calculator