We need a suitable gauge to use as a leak testing test set we are looking to build. We need to measure down to -10 mbar and monitor for 1 minute to ensure the leak rate is not greater than 1 mbar/minute.
We would suggest the Multicon CMC-99 device combined with a 10mbar vaccuum pressure transmitter.
The CMC-99 can be supplied as a panel mount device if you have your own panel, a freestanding benchtop mount chassis or a very simple wall mount enclosure. The CMC-99 can have an ethernet port so you can connect to your network or directly to a PC. You are provided with the free DAQ Manager software which can display the sensors in real time, whilst data logging for further analysis and downloading of data.
For leak testing purposes, the CMC-99 also has a timer function which you can use to count down the time, measure the first measurement, subtract the last measurement using the built-in Maths function, and then pass or fail the test visually on the display. The CMC-99 can also log the data between the start and stop function. The CMC-99 and vacuum sensor will be connected together and configured ready to plug & play.