Air Replacement Plant (ARP) 100 mm H2O differential pressure transmitter

Request Price

We are looking to improve efficiency of air replacement plant and would like technical data on differential pressure transmitters that we intend to incorporate before and after the ventilators of the insufflation of our paint booth air replacement plant, which we plan to monitor as part of an energy reduction project.   We want to use the differential pressure transmitters for instant monitoring of the air flow going through the painting booth interior.  The differential pressure readings will be in the order of 55 mm H2O and we will need a 4-20mA connection for the PLC (Programmable Logic controller) and an electronic display to indicate the measured value.

The DPS200 is a low range differential pressure transmitter specifically design for measuring ventilation air flow differential pressures. We would suggest  a 0 to 10 mbar range that is equivalent to around 100 mmH2O which would place your required pressure reading around the middle of the range.  The DPS200 also includes the option for a 4-20mA output and an integral LCD digital readout.

Air Replacement Plant (ARP) 100 mm H2O differential pressure transmitter

DPS200 Air Replacement Plant DP Transmitter

  • Pressure Range: 100 mmH2O diff
  • Output Signal: 4-20mA (2 wire)
  • Process Connection: 6.6 x 11 tube (standard)
  • LCD Display: Yes

Request Product Price

"*" indicates required fields

Please complete form below to request a quote for Air Replacement Plant (ARP) 100 mm H2O differential pressure transmitter.

Please enter info about your application and include any other requirements.

⚠ Attention Required

Please remove all web links from Application Details text entry box above, to enable form submission.

You will be able to provide any links related to your enquiry later on, by replying to the email confirmation once your submission has been processed.

Thank you.



Primary Sidebar

Pressure Sensor Specifier

Specify a pressure sensor

Select your pressure sensor requirements

Pressure Sensor Calculator

Calculate errors and readings for a pressure sensor
Enter your parameters into this pressure sensor calculator to determine calibration errors, expected readings, and error tolerances

Pressure Sensor Technology Guide

Rating table for performance of different pressure sensor technology types

Compare performance of the most popular pressure sensor technology types and how they rate for different application requirements

Wirelessly Measure & Log 4-20mA

Use this battery powered wireless 4-20mA wireless transmitter to transmit readings to a receiver connected to USB, RS485, RS232, cellular network or analog output

Use this battery powered 4-20mA wireless transmitter to transmit readings to a receiver connected to USB, RS485, RS232, cellular network or analog output

Products

  • Pressure
  • Level
  • Temperature
  • Interfacing
  • Datalogging
  • Calibration
  • Wireless

Product Enquiry

Send us your enquiry for Air Replacement Plant (ARP) 100 mm H2O differential pressure transmitter to get assistance with this product type.

Email Your Enquiry

enquiries@sensorsone.com
By clicking "OK" or continuing to use this website you are providing your consent to use cookies as described in our Privacy Policy

© 2025 SensorsONE Ltd, all rights reserved

  • Home
  • Products
  • Tools
  • Legal
  • Contact